Analytical Performance
Glucose Lactate Alcohol
Precision
(within run)
Plasma; n=20, CV=1.4%
@ 10.0mmol/L
Whole blood; n=40, CV=0.85%
@ 12.0mmol/L
Plasma; n=30, CV=0.6%
@ 10.0mmol/L
Whole blood; n=22, CV=1.6%
@ 5.0mmol/L
Plasma; CV=1-2%
@ 100mg/dl
Whole blood; CV=2-3%
@ 85mg/dl
Accuracy - Recovery Plasma; mean 100.9%
(99.2-106.7%)
range 1.5-30.0mmol/L
Whole blood; mean 96.7%
(89.1-100.2%)
Plasma; mean 99.2%
(96.0-104.0%)
range 2.5-10.0mmol/L
Whole blood; mean 96.7%
(88.0-104.0%)
Spiked serum;
y (Analox) = 0.973 (true)
+ 2.1mg/dl
r = 0.9929, n = 89
Accuracy - Method
Comparison, y = GM7
Plasma vs Hexokinase
y = 0.96 (HK) - 2.5mg/dl
r = 0.999; n = 156
Plasma vs Boehringer UV,
n = 20; r = 0.997
y = 0.967(UV) + 0.003
Whole blood vs YSI 23L,
n = 56, r = 0.998
y = 1.017(YSI) - 0.087
Whole blood vs GC,
y (Analox) = 1.04(GC)
+ 1.3mg/dl
r = 0.9905, n  = 27
Linearity 0 - 30mmol/L (540mg/dl)
using 10µl sample
0 - 50mmol/L (900mg/dl)
using 5µl sample
0 - 10mmol/L (90mg/dl)
using 7µl sample
0 - 20mmol/L (180mg/dl)
using 3.5µl sample
0 - 200mg/dl (43mmol/L)
using 5µl sample
0 - 400mg/dl (86mmol/L)
using 2.5µl sample

RETOUR