LABORATORY TECHNOLOGY

Belgique
France

Traduction

 

A premium, world class system, Analox Analysers offer unparalleled analytical performance and versatility to assist the Industrial, Commercial & Biotechnology markets. The Analox GL6 Analyser is an ultra-fast stand alone system for analysis of a range of chemistries in a range of fluids. The system offers exceptional precision and accuracy and is used in a variety of applications. It is delivered pre-programmed with 5 of your preferred single point calibration assays depending on you area of interest: Ethanol, Glucose, Glycerol, Lactate, Methanol, Sucrose or Lactose

Notes:

Example data below, please see Analyte Technical Bulletins for full specifications. The GL6 is supplied as standard with the 0.5ml reaction chamber.

 

 

 

Area Example Topics
Industrial Process Control  Using microbial oil produced from biodiesel by-products
Process control in monitoring the production of dairy products
Food & Beverage Production As a marker for spoilage in eggs, milk, tinned fruits and vegetables
Improving the flavour and stability of foods
Detecting methanol contamination in beer, spirits and wine
Fermentations in Food & Beverage Production Monitoring malolactic fermentation in wine (Oenology)
Pharmaceuticals Developing drugs for sepsis
Humectant use in drug development
Biotechnology Increasing the sucrose content of sugarcane for production of chemicals, medicines and biofuels
Offshore Methane Fuel Extraction Developing sustainable gasoline blending options
Toiletry Production Development of oral care/hygiene products

 

After calibration with an appropriate standard, the single injection of sample is all that is needed to obtain a result and prepare the analyser for the next analysis.

Sample injection via an accurate positive displacement pipette triggers the complete analytical cycle and a hard-copy result is printed within 25 seconds

 

Precision Accuracy (Method Comparisons) Linearity
Glucose 0.01 g/L @2g/L N/A 0.05 - 40 %W/V
Lactate 0.01 %W/V N/A 0.06 – 14 mmol/L (ca. 0.5 - 126 mg/dl)
Ethanol 0.01 %V/V, 0.001% (selectable) N/A 0.1 – 10 %V/V
Glycerol 0.01 %W/V N/A 0.1 - 4.0 %W/V (1.0 - 40.0 g/L)
Methanol 0.01 %W/V N/A ca. 5 - 500 ppm ca. 0.05 - 4 %W/V (aqueous dilution)

 

 

 

 

Instrument Specification:
Method Enzymatic oxygen-rate
Sensor Clark-type amperometric oxygen electrode
Reaction Temperature 30°C
Display 32 character backlit LCD
Printer 16 column dot matrix, 1 line/sec.
Statistics Sequential, giving Mean, S.D and C.V.
Interface Serial data port, optional Windows software available
Power 100-250VAC, 50-60Hz, 12-15VDC, 60VA
Dimensions 23cm(width) x 29cm(depth) x 15cm(height)
Weight  3.8Kg 

 

 

 

 

 

 

 

 

 

 

 

IMLAB bvba
Oude Vijvers 1
B - 3370 Boutersem
Belgique

Tél.:  016 - 73 55 72

IMLAB sarl
46 rue des Canonniers
F - 59000 Lille
France

Tél.:  03 20 55 19 11